LIGO Production Large Actuator Test Report |
Actuator Serial #: L021
Operator Name: Smith, Lane
Date: 8/12/2009 Time: 11:38 AM
Actuator Coil Resistance: 6.36 Ohms, PASS
Ambient Temperature: 70.4 F
Hi Pot Test Results: 1000 MOhms, PASS
X Travel Limit (inches): 0.523
Y Travel Limit (inches): 0.206
Z Travel Limit (inches): 0.507
|
X Position (in) |
Y Position (in) |
Z Position (in) |
Current (A) |
Force (lbs) |
Force Constant (lbs/A) |
PASS/FAIL |
Center |
0.000 |
0.000 |
0.000 |
2.146 |
20.332 |
9.474 |
PASS |
Position 1 |
0.250 |
0.075 |
0.215 |
2.442 |
19.910 |
8.153 |
PASS |
Position 2 |
0.250 |
-0.075 |
0.215 |
2.366 |
19.959 |
8.436 |
PASS |
Position 3 |
-0.250 |
-0.075 |
0.215 |
2.339 |
20.020 |
8.559 |
PASS |
Position 4 |
-0.250 |
0.075 |
0.215 |
2.358 |
20.183 |
8.558 |
PASS |
Position 5 |
-0.250 |
0.075 |
-0.215 |
2.366 |
20.174 |
8.528 |
PASS |
Position 6 |
-0.250 |
-0.075 |
-0.215 |
2.338 |
19.897 |
8.511 |
PASS |
Position 7 |
0.250 |
-0.075 |
-0.215 |
2.283 |
19.901 |
8.717 |
PASS |
Position 8 |
0.250 |
0.075 |
-0.215 |
2.347 |
20.214 |
8.612 |
PASS |
Center |
0.000 |
0.000 |
0.000 |
-2.075 |
-19.894 |
9.587 |
PASS |
Position 1 |
0.250 |
0.075 |
0.215 |
-2.335 |
-20.164 |
8.637 |
PASS |
Position 2 |
0.250 |
-0.075 |
0.215 |
-2.314 |
-20.102 |
8.688 |
PASS |
Position 3 |
-0.250 |
-0.075 |
0.215 |
-2.281 |
-19.955 |
8.750 |
PASS |
Position 4 |
-0.250 |
0.075 |
0.215 |
-2.320 |
-20.067 |
8.649 |
PASS |
Position 5 |
-0.250 |
0.075 |
-0.215 |
-2.430 |
-19.865 |
8.175 |
PASS |
Position 6 |
-0.250 |
-0.075 |
-0.215 |
-2.496 |
-19.850 |
7.952 |
PASS |
Position 7 |
0.250 |
-0.075 |
-0.215 |
-2.484 |
-20.105 |
8.093 |
PASS |
Position 8 |
0.250 |
0.075 |
-0.215 |
-2.370 |
-19.970 |
8.427 |
PASS |